Product information
The graph below illustrates the maximum amount of solvent allowed for each load at a given working temperature.
Solvent with molar mass M = 100g/mol
Minimum explosion limit U = 40g/m at 20°C
Safety factor K = 0.5
Example
Solvent used, type mentioned above
Room temperature +20°C
Operating temperature +80°C
Maximum amount of solvent allowed 11g
• No. 02 Removable and height-adjustable shelves
• N°01 Power cable
• N°01 Instruction manual
• N°01 CE Certificate of Conformity
• N°01 Certificate of Guarantee
• AG-System: Digital PID microprocessor control, accurate and sensitive. AGS Control DAILY, PRO and TALENT versions
• Temperature range: from room temperature +10.0°C to +250°C
• Safety Thermostat, Class 3.1
• Suitable for drying samples containing solvents
• Complies with EN 1539
• Reproducibility of the tests performed: in applications where the formation of an explosive mixture inside the chamber must be avoided, for example in the use of paints, lacquers, adhesives, binders, impregnated coils
• Forced Air circulation system, dynamic and homogeneous
• Made in Italy